Groupe Européen de Recherche sur les Prothèses appliquées à la Chirurgie Vasculaire, Strasbourg, FranceDepartment of Vascular Surgery and Renal Transplantation, Nouvel Hôpital Civil, Strasbourg, France
Groupe Européen de Recherche sur les Prothèses appliquées à la Chirurgie Vasculaire, Strasbourg, FranceDepartment of Vascular Surgery and Renal Transplantation, Nouvel Hôpital Civil, Strasbourg, France
Groupe Européen de Recherche sur les Prothèses appliquées à la Chirurgie Vasculaire, Strasbourg, FranceDepartment of Vascular Surgery, Bedford Hospital NHS Trust, Bedford, UK
Corresponding author. Department of Vascular Surgery and Kidney Transplantation, Les Hôpitaux Universitaires de Strasbourg, B.P. 426, 67091 Strasbourg Cedex, France.
Groupe Européen de Recherche sur les Prothèses appliquées à la Chirurgie Vasculaire, Strasbourg, FranceDepartment of Vascular Surgery and Renal Transplantation, Nouvel Hôpital Civil, Strasbourg, FranceDepartment of Physiology and Respiratory Functional Explorations, Nouvel Hôpital Civil, Strasbourg, France
The impact of endoanchors penetration on endograft textile structure has been studied.
•
Systematic analysis of two aortic explanted endografts was performed.
•
Textile alteration, tears of the binding threads and stents lesions were observed.
•
Site of penetration of endoanchors might contribute to endograft fabric damage.
•
Follow-up on new technologies through explant analysis programs is mandatory.
Objective
The adjunctive use of Heli-FX EndoAnchors (EAs) in endovascular aneurysm repair has been proposed for the treatment or prevention of type Ia endoleaks. The aim of this study was to evaluate the impact of the penetration of EAs on endograft textile structure from two aortic endografts that had been explanted for persistent type Ia endoleak despite the implantation of EAs.
Methods
An Aorfix aortic endograft was implanted in an 85 year old man. Six months later, six EAs were implanted for Ia endoleak. The endograft was explanted as an emergency seven months later for aneurysm rupture. An Endurant II aortic endograft was implanted in an 80 year old man. Seven EAs were implanted 24 months later for type Ia endoleak. A proximal cuff extension with bilateral renal and superior mesenteric artery chimneys was performed 18 months later to treat a persistent type Ia endoleak. Endograft explantation was performed six months later owing to persistent type Ia endoleak and aneurysm sac enlargement. Explant analysis in both cases was performed at GEPROVAS.
Results
Systematic analysis of both explants, including the 13 EAs, revealed the following lesions: (1) alteration of textile structure directly linked to several penetrations of the fabric with the same EA and tears of the textile fibres in two cases; (2) tears of the binding threads as the EA had passed through them in five cases; and (3) interactions between EA and endograft stents in four cases.
Conclusion
The site of EA penetration into the endograft might contribute to endograft fabric damage and to a loss of stability of the endograft at the level of the aortic neck.
Endovascular aortic aneurysm repair efficacy is known to be significantly linked to proximal sealing and fixation. This usually implies a sufficient neck length in order to avoid type Ia endoleak. Instructions for use (IFU) recommend at least 15 mm for most devices, although it can be decreased to 10 mm for some endograft models, such as the Endurant (Medtronic, Santa Rosa, CA, USA) or the Incraft (Cordis, Baar, Switzerland) devices. Recently, the Endurant device IFU included primary adjunctive deployment of Heli-FX EndoAnchors (EAs) for short necks (4–10 mm), in order to obtain better device fixation.
However, the rate of type Ia endoleak after secondary fixation remains about 22.6% and adverse events include EA fractures, EA dislocation, or EA entrapment.
Moreover, current evidence lacks long term follow up and perhaps case controlled trials are recommended for EA use in routine practice, although this is already integrated in some.
As the principle of the EA technique is penetration of the endograft structure with a screw, it seems right to assess the risk of a potential deleterious impact on the materials, in order to identify factors that could influence the long term stability of the devices that might lead to a failure or complications.
The aim of this study was to evaluate the impact of EA penetration on endograft textile structure from two aortic endografts that had been explanted for persistent type Ia endoleak, despite the implantation of EAs.
Reports
The first case was of a 24 mm diameter Aorfix (Lombard, Didcot, UK) aortic endograft that was implanted in an 85 year old man for a 75 mm diameter abdominal aortic aneurysm (AAA) with 60° neck angulation. Computed tomography angiography (CTA) at six months revealed a type Ia endoleak with an increase in aneurysm sac diameter due to endograft proximal malapposition. Six EAs were implanted. Completion angiography did not show any residual type Ia endoleak. Seven months later, the patient presented with AAA rupture. The aortic wall was cut at the margins of the EAS and explantation of the proximal part of the endograft followed by an aortobi-iliac prosthetic bypass was performed as an emergency. The post-operative course was uneventful.
The second case was of a 25 mm proximal diameter Endurant II aortic endograft implanted in an 80 year old man with a 61 mm diameter AAA. Follow up CTA at 24 months revealed a type Ia endoleak with an increase in aneurysm sac diameter due to endograft malapposition. Seven EAs were implanted. Follow up CTA at 18 months showed recurrence of a type Ia endoleak with an increase in aneurysm sac diameter. A proximal cuff extension with bilateral renal and superior mesenteric artery chimneys was performed. Follow up CTA at six months showed persistence of the type Ia endoleak, with continued significant aneurysm sac enlargement. The aortic wall was cut at the margins of the EAs and explantation of the proximal part of the endograft with visceral arterial revascularisation and aorto-aortic bypass was performed. The post-operative course was uneventful.
After explantation, both specimens were sent to GEPROVAS, as a part of a European retrieval programme.
Results
After explantation, the specimens were submitted to a standardised protocol for evaluation (Supplementary Material) starting with naked eye examination followed by digital image capture (Nikon D5100, Nikon France, Champigny sur Marne, France). The specimens were then cleaned using a papaine based enzymatic cleaning process at 60°C, allowing complete digestion of the aortic wall. After cleaning, the specimens were rinsed using distilled water and additional pictures were taken with a Keyence VHX 600 digital microscope (Keyence France, Courbevoie, France).
The first case consisted in the bifurcated body of an endograft where six EAs had been implanted (Fig. 1A). Each EA was seen on the endograft and named from EA1 to EA6. The second case consisted of the bifurcated body of an endograft where seven EAs had been implanted (Fig. 1B). Only four EAs were seen on the endograft; each EA was named from EA7 to EA10. Three EAs were detached during the cleaning process and the visible holes were designated H1, H2, and H3.
Figure 1(A) First explant after cleaning, showing six EndoAnchors (EA; named EA1 – EA6) in place. (B) Second explant after cleaning, showing four EAs (named EA7 – EA10) in place and three holes (H1 – H3), as three EAs came off during the cleaning process.
Observations made on each EA are reported in Table 1. Three different types of interactions were observed: between EA and endograft fabric; between EA and ligature threads fixing the stents; and between EA and stents. Concerning EA–endograft fabric interactions, eight of the 10 EAs penetrated the fabric only once, while two EAs passed three and five times, respectively, through the fabric, because there were three and five holes, respectively, in the vicinity of the same single EA. Some fibre breakages and opening of the textile structure were observed (Fig. 2A). Concerning the interaction of the EAs and ligatures, four EAs passed through the stitches, creating thread dissociations and filament ruptures (Fig. 2B). Concerning EA–stent interactions, in three of the 10 cases the EAs penetrated close to the stent; in one case the EA made close contact with the external part of the stent and in one case an EA linked the suprarenal bare stent to the first stent of the first covered stent of the endograft (Fig. 2C). In this case, major wearing lesions on both the EAs and endograft stent were observed.
Table 1Macroscopic view (50×) of each EndoAnchor (EA; named EA1 – EA10); two holes (H1 and H2) left by EA detached during the cleaning process along with the characteristics of their path through the endograft and one lesion (L1).
Picture
Number of penetrations
Angle of penetration
Relation with stitches
Relation with stent
EA1
1
630°
Passed through a thread
No contact
EA2
5
540°
No contact
No contact
EA3
1
540°
Passed through one thread
Penetrated close to the stent and in close contact on its external side
EA4
1
540°
No contact
No contact
EA5
3
810°
Passed through two thread
Linked to the stent by the suture thread
EA6
1
540°
No contact
No contact
EA7
1
270°
No contact
No contact
EA8
1
540°
Passed through one thread
Penetrated at the contact of the stent and close connection with it
EA9
1
630°
No contact
In close contact to the suprarenal stent and first row of the endograft stent, signs of wearing on both stents and EA
Figure 2(A) Multiple passes of the EndoAnchor (EA; EA2) through the fabric causing alteration of the textile structure (Keyence magnification 50×). (B) Partial rupture of a textile mesh due to the EA passing (EA3) through the endograft stitches (Keyence magnification 150×). (C) EA9 passing close to the suprarenal stent (Keyence magnification 50×). (D) Mesh enlargement and partial rupture of the fibres on hole (H2; Keyence magnification 150×).
The three holes identified after cleaning demonstrated an enlargement of the textile structure and some partial rupture of the fibres, one of which showed alteration of a line of stitches (Fig. 2D).
Discussion
To the authors' knowledge this is the first report on explanted endografts, including EAs, and it was seen that the use of an EA could alter the structure of the endograft in different ways.
In the majority of cases, the EA penetrated the fabric far from the stents and the stitches with only one fabric penetration. The penetration, evaluated in degrees, showed variable penetration up to 810°. However, that the explantation could have modified this degree of penetration cannot be excluded. Interestingly, in two cases the EA passed three and five times through the fabric, respectively, meaning that the screw passed first through the fabric and did not immediately penetrate the arterial wall but re-entered the fabric from outside (once in one case and twice in the other case) before penetrating the arterial wall. This can potentially be explained by several mechanisms: slipping of the EA on the graft because of a lack of strength during the apposition of the delivery system; because of difficulties penetrating the arterial wall related to local factors such as calcification; or multiple attempts at deploying the EAs in the same area, as can happen.
However, when the EA was inserted close to particularly important structures, such as the stents and the ligature threads, it was found that these interactions could lead to structural lesions of the threads with rupture of filaments and of the stents with wearing lesions. This is particularly true when the EA fixed the two crests of the stents together. Moreover, these interactions could lead to a concentration of stresses on areas that were initially independently mobile and able to absorb initial mobility that become closely linked. However, the long term effect of this cannot be hypothesised as expert users are aware that EAs sometimes brush past stent metal during deployment.
The EA IFU recommends distributing the EA implants as regularly as possible around the circumference of the sealing stent,
Even though the delivery system allows accurate placement, it may be difficult to know where exactly the tip of the screw will engage in terms of fabric and stent metal. Consequently, the risk of penetrating the endograft structure close to the ligature threads and/or the stents is not negligible and can sometimes immediately result in maldeployment issues such as migration and on table EA fracture.
The observations showed that deleterious interactions of the EA could occur in some instances and both cases were explanted for type Ia endoleaks. It does not necessarily mean that these failures were directly related to deleterious interactions.
Following studies on endograft displacement forces after endostapling,
These results directly suggest that good positioning and deployment of the anchor requires specialised skills and a trained team, with experts suggesting a multiplanar approach for hostile necks.
However, the design of these studies did not integrate the long term cyclic load, but focused on traction resistance that results in exposure to forces much higher than in vivo forces and did not integrate the arterial cyclic load, which created the lesions observed. However, it is worth considering that optimal EA deployment achieves good aorto–graft apposition, reinforcing the suggestion of the protective effect of EAs even with continued aortic neck dilatation,
which may be an effect of both static (radial force) and cyclic (haemodynamic) loading.
This suggests that even if EAs might be promising treatment for some patients, they should be used carefully and strictly following the manufacturer's IFU. In the first case, EAs 4, 5, and 6 were deployed too low and were probably not engaging the aorta at all. The principle of circumferential EA deployment was not followed. EA deployment needs to be accurately planned out, because “blind” placement might lead to persistent complications requiring complex open revisional surgery in high risk patients.
Editor's choice – European Society for Vascular Surgery (ESVS) 2019 clinical practice guidelines on the management of abdominal aorto-iliac artery aneurysms.
The main interest of a retrieval programme of explanted prostheses is to allow the analysis of a significant number of prostheses, in order to describe the ageing characteristics of the different models of devices and to learn about the concept itself. However, a study based on a retrieval programme has unavoidable limitations that must be taken into consideration when interpreting the results as one may consider that the probability of finding structural lesions on endografts explanted for a complication is higher than on uncomplicated endografts. Moreover, that lesions could have been created during the explantation of the endograft and therefore are a result of manipulation rather than damage resulting from EA deployment and fixation must be considered.
The ideal specimen for analysis would result from the autopsy of a patient who died of unrelated causes. Finally, the time between EA placement and explantation is short. Further studies are therefore needed in order to assess the long term effects of EAs on textile.
Conclusion
This study confirms the necessity of following up new technologies through explant analysis programmes and provides insights on the impact of EAs on endografts, depending on where the EAs enter through the fabric. Close follow up of patients treated with EAs remains necessary and further studies focusing on these interactions under cyclic load are still required.
Funding
The authors are indebted to the Eurometropole de Strasbourg and the Région Grand’Est for their financial support.
Conflict of interest
None.
Acknowledgments
The authors acknowledge the European Society of Vascular Surgery and the Société Francaise de Chirurgie Vasculaire et Endovasculaire, who support the explant analysis program.
Appendix A. Supplementary data
The following is the supplementary data related to this article:
Editor's choice – European Society for Vascular Surgery (ESVS) 2019 clinical practice guidelines on the management of abdominal aorto-iliac artery aneurysms.
To submit a comment for a journal article, please use the space above and note the following:
We will review submitted comments as soon as possible, striving for within two business days.
This forum is intended for constructive dialogue. Comments that are commercial or promotional in nature, pertain to specific medical cases, are not relevant to the article for which they have been submitted, or are otherwise inappropriate will not be posted.
We require that commenters identify themselves with names and affiliations.